geko™ T-1

The wrist-watch sized geko™ device is a disposable, internally battery powered, neuromuscular electrostimulation unit. It is an integrated device and there are no accessories required for its operation. Its intended use is:

  • to increase venous circulation, and
  • for the prevention of venous thrombosis

 

Available for use in Europe, Australia, New Zealand and Canada for increased venous circulation and for the prevention of venous thrombosis, the geko™ T-1 device is not approved by the US FDA and is not available for sale in the USA.

 

Specifications:

product name geko
model reference T-1
product type powered muscle stimulator
class BF
dimensions 149mm x 42mm x 11mm
weight 18g (geko™ device only)
power source internally powered equipment, battery not replaceable
battery primary lithium coin cell, removable for disposal
operation

continuous operation - equipment not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide

stimulation modes 7 (selectable pulse widths)
pulse current 27mA (±15%) constant current, compliance to 80v
load impedance 200Ω to 3kΩ for 27mA output
pulse voltage set by current and load
pulses width
(200Ω to 10kΩ)
70, 100, 140, 200, 280, 400 and 560µs (±5% +20µs)
(open circuit ±5% +70µs)
repetition rate 1Hz (±5%)
maximum charge 20µC per pulse
net charge output zero per cycle
output coupling ceramic capacitor

 

 

 

 

 

1397T-1 spec_wave_L

click-image-enlarge-dots


1395T-1 spec_output_L

click-image-enlarge-dots

 

current rms (500Ω) 1mA rms maximum
voltage rms (500Ω) 0.5V rms maximum
operating time 24 hour duration (maximum 30 hours)
indicator display green LED, flashing
fault indication the stimulator device will automatically switch off for over current, under current, low battery voltage or end of 30 hour run time
materials soft moulding: TPE
Hydrogel cover: PET (Mylar) polyethylene terephthalate
standards IEC60601-1 (1996)
IEC60601-2-10 (2001)
IEC60601-1-2 (2004)
ISO 10993

 

Operating conditions:

temperature range 10°C to 30°C
humidity range 10% to 80% non-condensing

 

Storage conditions (in original packaging):

temperature range -10ºC to 30ºC
humidity range 10% to 80% non-condensing
shelf-life 24 months when stored as above


Warranty:

24 months storage before use and 24 hours stimulation

Downloads
  • geko™ T-1 fitting instructionsPDF documentDownload button
  • geko™ T-1 instructions for usePDF documentDownload button